Type TLX-120AP1 TLX-200AP

Measurement range 0–12 m 0–20 m

Measurement method Pulse method

Measurement scale Wire rope Φ2mm, SUS304

Motor Single-phase 20 W reversible (50/60Hz), Electromagnetic brake type

Current output 4–20 mA DC, resistance load 350 Ω or less

Pulse output NPN open-collector output

Start output Maximum 40 mA (30 A) or less, residual voltage 1 V or less

Output during measurement Contact a, AC250V/5A, DC30V/5A (resistance load) Output

Abnormality output Contact a, AC250V/5A, DC30V/5A (resistance load)

Input Non-voltage contact signal (start input)

Measurement resolution 10 cm

Accuracy 10 cm 1% of F.S.

Indicator lamp: Power supply (yellow), operation (with a two-color LED, measuring: green, abnormality: red)

Weight elevating speed About 13 m/min

Timer function 3 min, 5 min, 10 min, 15 min, 20 min, 30 min, 60 min, 2 h, and 3 h (3 min cannot be selected with TLX-200AP))

Power supply voltage AC100V or AC200V, 50/60Hz

Power consumption 70 VA

Flange size 5K100A (SS400)

Ambient temperature during use -10ºC to +50°C (no freezing allowed)

Materials used (main) Body, flange SS400 Cover SS400 Wire SUS304 Drum Cast aluminum Weight (standard) SS41

Painting color Munsell 10YR7.5/14

Weight 31 kg